Measuring Instruments
Data Logger
Data Logger
   EDX-5000A is an advanced all-in-one measuring instrument having sophisticated features and high-speed processing capabilities.It is the high-end model of EDX Series.
Product Details

Model

EDX-5000A-64-HE: Hard disk drive (HDD) , English
EDX-5000A-64-SE: Solid state drive (SSD) , English

Applicable Conditioner Cards

CVM-41A, CDV-40B(-F), DPM-42B(-F), DPM-42B-I(-F),
CTA-40A, CFV-40A, CCA-40A(-F), CDV-44AS, CDA-44AS, CDA-45AS, AD-40AS(-F)

Input Channels

Max. 64 (CDV-40B x 8)

Analog Input

See specifications of respective conditioner cards for details.

Digital I/O

I/O points: Max .8 
I/O setting: Sets every bit to digital input, digital output, or remote-control input with the same common. 
*Remote-control input: Measuring start, stop, balance conducting, etc. 
Input modes: Isolated TTL Level input 
Input voltage: Max. 5 VDC 
Isolation modes: Digital isolation 
Output modes: Isolated open-collector output with 10 kohm built-in pull-up resistors 
Output voltage: 5 VDC 
Output current: Max. 25 mA per point

Voice Input

1 channel, to be recorded together with measuring data. An optional remote control unit RCU-42 is required.

Sampling

Sampling Frequencies : 1-2-5 series
1 Hz to 200 kHz for up to 32-channel data acquisition
1 Hz to 100 kHz for up to 64-channel data acquisition
1 Hz to 50 kHz for up to 80-channel data acquisition
1 Hz to 10 kHz for real-time synchronous data processing
2n series
2 to 131072 Hz for up to 32-channel data acquisition
2 to 65536 Hz for up to 64-channel data acquisition
2 to 32768 Hz for up to 80-channel data acquisition
2 to 8192 Hz for real-time synchronous data processing
The condition is using a 16-bit A/D converter, as for a 24-bit A/D converter, the channels above will be half of each channel.

Storage

HDD model: 300 GB or more
SSD model: 30 GB or more

Display

Channel status LED (OVER value of every channel are settable), REC, BUSY, BATTERY, POWER LED
12.1" wide touch panel

Operation Keys

Keys REC, STOP, BAL, and OPT are operated by touch panel,
keys on EDX, external keyboard, or mouse

Output Control connectors

CONT.IN, CONT.OUT (For remote control unit, or synchronous operation)

External I/O Connectors

External Triggers: TRG IN, TRG OUT
External clock: CLK IN, CLK OUT
(Output at any frequency division ratio)

External Interfaces

USB type keyboard, USB type mouse
VGA connector for monitor
USB I/F: 3 USB 2.0 in front, 2 USB 3.0 in rear
LAN I/F: 100/1000BASE-T

Power Supply

100 to 240 VAC and 10 to 30 DC, dual supply
Built-in batteries against momentary power failure

Current Consumption

Approx. 1.8 A (For 100 VAC, CDV-40B x 8)

Operating Temperature

0 to 40 ºC

Operating Humidity

20 to 80% RH (Non-condensing)

Storage Temperature

-20 to 60 ºC

Vibration Resistant

EDX-5000A-64-H, EDX-5000A-64-S,
49.0 m/s2 (5 G), 5 to 55 Hz when not operating
29.4 m/s2 (3 G), 5 to 55 Hz when operating
EDX-5000A-64-H: 
9.8 m/s2 (1 G), 10 to 200 Hz when operating
EDX-5000A-64-S: 
19.6 m/s2 (2 G), 10 to 200 Hz when operating

Impact Resistant

196.1 m/s2 (20 G), 11 ms

Dimensions

365 W × 159 H × 300 D mm (Excluding protrusions)

Weight

Approx. 11.5 kg (Mainframe only)

Optional Functions

CAN data acquisition (Input channels: Max 512)
GPS data acquisition (Tracing data acquisition,
time synchronous function)

Online Control

Controlled by DCS-100A

EMC Directive

EN61326-1 (Class A)

Low Voltage Directive

EN61010-1(Installation category II, Pollution degree 2)

RoHS Directive

EN50581