Data Logger
Data Logger
Data Logger

   The UCAM-60B is an all-in-one Data Logger, which has easy to operate keys, easy to visual display and a printer for immediate confirmation of measurement results.


Product Details

Types

UCAM-60B-AC AC only
UCAM-60B-DC DC only

TEDS

Interface: IEEE 1451.4 Mixed Mode Transducer Interface Class 2
Applicable sensor: Should have information written in accord-ance
with IEEE template No. 33; cable length should be 30 m or less

Scanning Speed

Notes: 
1. High-resolution mode and high-speed mode are selectable for dedicated scanners only.
2. High-resolution or high-speed mode is available only with full bridge system.

File Conversion

CSV conversion

Coefficient Calculation

Multiplication by calibration coefficients, calibration by TEDS, conversion of measured values to physical quantities, scaling and correction.

Self-diagnosis Functions

Checks display, printer, bridge excitation, lead-wire-off, input/output
resistance, insulation resistance, mode, etc. Checking of input/outut
resistance and mode is availabe only for dedicated scanners.

Automatic Measurement Functions (Trigger Measurement)

A relative value (Certain changing quantity) or an absolute value
triggers measurement. In addition to the usual trigger function,
a variable trigger function is provided with which the trigger value
changes at each step during measurement. With this special function, a trigger value and the number of measurement times (Repeat times) under the trigger condition can be registered for each step to perform a series of automatic measurements in the order of steps.

Automatic Measurement Functions (Interval MeASUREMENT: )

Measurement is automatically performed at preset time intervals.

Automatic Measurement Functions (Trigger Interval Measurement)

Combination of trigger measurement and interval measurement.

Channels

Max. 30 with dedicated scanners
Max. 1000 with external scanners connected

Measuring Functions

Initial: Initial values are measured and stored in internal memory
(Except for temperatures measured by civil engineering transducers with temperature measuring function).
Original: Raw values are measured without subtraction of initial values.
Measure: Initial values are subtracted from original values (Except for temperatures measured by civil engineering transducers with temperature measuring function).
Easy Measure: Auto zero balancing function is activated.
Note: The selected function is applied to all channels. Coefficient

Measuring Targetss

Strain gages, strain gage transducers, civil engineering transducers
with temperature measuring function, DC voltage-output or DC current-output instruments, potentiometer sensors, thermal sensors (Thermocouples and platinum resistance thermometer bulbs)

Units

59 units

Operation Modes

Real-time, Monitor, and Automatic

Standard Accessories

AC power cable P-18 with 2-pin conversion plug CM-39 (AC-operated version)
DC power cable P-76 (DC-operated version)
Recording paper UCAM-60A-RP (1 roll for UCAM-60B only)
Screwdriver
Spare fuse
CD-R (Instruction Manual)

Optional

Recording Paper UCAM-60A-RP (10 rolls/pack)

Sacnning Speed

USB-70B (Standard Mode only) : Standard Mode (50 ms/channel) Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s

Auto Measuring Functions (Trigger Measurement)

Trigger channel : 1 desired channel

Automatic Measurement Functions (Trigger Measurement)

Trigger Value : A desired real number of 6 effective figures or less Reference

Auto Measuring Functions (Trigger Measurement)

standard Value : Amount of level shift to determine the first trigger value (Selected from the same range as for the trigger values)
Repeat Times : 1 to 9999 (0 for infinite times)
Measuring Steps : Maximum 15

Auto Measuring Functions (Interval Measurement)

Measuring Start Time : Year/month/day and hour: minute: second
Interval setting: Days and hours: minutes: seconds (In a range of 00 and 00: 00: 01 to 99 and 12: 59: 59)
Measuring Interval : Days and hours: minutes: seconds (In a range of 00 and 00: 00: 01 to 99 and 12: 59: 59)
Repeat Times : 1 to 9999 (0 for infinite times)
Available Steps : Maximum 15

Auto Measuring Functions (Trigger Interval Measurement)

Measuring Times : Max. 9999
Interval Time : Available in a range of 1 second to 99 days 23 hours: 59 minutes: 59 seconds

Storage Functions

Built-in Memory : Internal memory, approx. 7 MB
Exterior Memory : Flash ATA card (Optional); the capacity depends on the card.

Sacnning Speed

Exclusive Scanning Unit (Standard Mode): : 50 ms/channel. Scanning speeds stated above are standard maximum speeds in respective modes. 
Besides these, the following speeds can be set for each individual channel: 0.28 s, 0.5 s, 1 s, 2 s, 5 s and 10 s
Exclusive Scanning Unit (High-Conversion Mode): : 280 ms/channel (High-resolution mode)
Note: Individually selectable for desired channels
Exclusive Scanning Unit (High-speed Mode): : 20 ms/channel (High-speed mode), voltage, Current, only Potentiometer sensors

Top

Copyright © 2017 Shenzhen Qualitykon Technology Co.,Ltd. All Right Reserved 

Tel: 0755-81464182    Fax: 0755-89370613     Address: No. 319, No. 258, Xingye 1st Road, Rentian community, Fuhai street, Bao'an District, Shenzhen